LK-H008 Laser Displacement Sensor
Model
LK-H008
Mounting Method
Mirror Reflection
Reference Distance
8 mm / 0.315"
Measurement Range
±0.5 mm / ±0.02"
- *1
Light Source
- Type: Semiconductor Red Laser
- Wavelength: 655nm
- Laser Class (JIS C6802): Class 1
- Emission: 0.3mW
- Spot Diameter (at Standard Distance): ø20 µm
Linearity
±0.05% of E.T. (E.T.= 1 mm / 0.04")
- *2
Repeatability
0.005 µm (0.001 µm)
- *3
Sampling Cycle
2.55/5/10/20/50/100/200/500/1000 µs (9 selectable levels)
Temperature Characteristics
0.02% of E.T./°C (E.T.= 1 mm / 0.04")
Environmental Resistance
- Degree of Protection: IP67
- Ambient Light: Incandescent lamp or fluorescent lamp: 10,000 lux max.
- Ambient Temperature: 0 to +50 °C / 32 to 122 °F
- *4
- Relative Humidity: 35 to 85% HR (No condensation)
- Vibration Resistance: 10 to 55 Hz, Double Amplitude 1.5 mm / 0.06", 2 hours in each of the X, Y and Z directions
Material
Stamped Aluminum
Weight
Approx. 240 g
- *1 Range of measurement when the sampling cycle is 20 µs or more.
- *2 This value is obtained when the KEYENCE standard object (diffused white piece or, for the LK-H008/LK-H008W only, metallic piece with specular surface) is measured in normal measurement mode.
- *3 This value is obtained when the KEYENCE standard object (diffused white piece or, for the LK-H008/LK-H008W only, metallic piece with specular surface) is measured at the reference distance with the average number of measurements set to 16384. The value in parentheses is a typical example of a measurement with the average number of measurements set to 65536 and the sampling cycle at 200 µs.
- *4 When the ambient temperature rises to 40 °C / 104 °F or more, mount it on a metal plate before use.