Calibrador de termoelementos PCE-TTC 30

Calibrador de termoelementos PCE-TTC 30

  • Code / Model:
  • Manufacturer: PCE Instruments
  • Department:
  • Category of products:
  • Sub-category: Instruments and equipment
  • You've seen it before.: 1
Heat element calibrator for simulation and measurement of heat elements, current and voltage / Li-Ion accumulator / Continuity testing / Ramp function
The thermoelements calibrator is designed to simulate different types of thermoelements and mV. This is why it can use the thermoelements calibrator for different thermometers and temperature indicators with connection to thermoelements.

In addition to introducing a temperature or voltage value up to 250 mV, the thermoelectric calibrator can also be used in ramp mode. The user can decide whether to use the ramp in continuous mode or at defined intervals.

- Simulation and measurement mode
- The accumulator
- For thermometers E, J, K, T, B, R, S, N, mV
- 24 V power supply for loop
- Compatible with HART
- Manual mode and ramp function
- Continuity test
- Data recording function

The following shall be added:

Rate

0 ... 30 V

Resolution adopted by the Council

0,001 V

Accuracy

±0,02 % of the value + 2 dgt.

 

Other electrical equipment

Rate

0 ... 24 mA

Resolution adopted by the Council

0,001 mA

Accuracy

±0,02 % of the value + 2 dgt.

 

Other heat-treating

The guy

Rate

Resolution adopted by the Council

Accuracy

And then

-200 to 1200 °C

The following conditions shall apply:

The following conditions shall apply:

J

-200 to 1200 °C

The following conditions shall apply:

The following conditions shall apply:

K

-200 ... 1372 °C

The following conditions shall apply:

The following conditions shall apply:

T

-200 ... 400 °C

The following conditions shall apply:

The following conditions shall apply:

B. Other

450 to 1800 degrees

The following conditions shall apply:

The following conditions shall apply:

R

0 ... 1750 °C

The following conditions shall apply:

The following conditions shall apply:

S

0 ... 1750 °C

The following conditions shall apply:

The following conditions shall apply:

N

-200 to 1300 °C

The following conditions shall apply:

The following conditions shall apply:

mV

-10 ... 80 mV

0,001 mV

± 0,02 % of the value + 2μV

 

-10 ... 250 mV

0,01 mV

± 0,02 % value + 0,02 mV

 

Note:These indications correspond to the ITS-90 temperature standard.

 

Other specifications

Screen modes

Measurement: mA/mV/V/thermoelectrical
Simulation: mV / thermoelectric

Heat and air

The following is the list of the following:

Error of cold joint compensation

Other, of a kind used for the manufacture of goods

Maximum input voltage

30 V DC

The temperature coefficient

Other, of a thickness of < 30 ppm

Impedance of entry

The heat element / mV / V: > 1 MΩ
Measurement of current: 10 Ω

Response time

Other, of a kind used for the manufacture of goods

Impedance

> 4,7 kΩ with thermoelectric power / mV

Frequency of display updates

10 Hz

Isolation

Other, of a kind used for the manufacture of goods

Data recording

Internal memory, 150000 values

Interface

Other electrical equipment

Screen

TFT LCD 2,4" / 240 x 320 pixels / Background lighting

The output voltage of the loop shall be:

The following conditions shall apply:

The following shall be added to the list of the following:

250 Ω ± 20%

Special functions

Passage and ramp function
Manual and automatic mode
√x, x2: For the measurement function

Continuity test

Adjustment of the limit value up to 100 Ω

Food

Li-Ion battery with a capacity of 3,7 V / 2300 mAh

Duration of load

About five hours.

Source of power

Input: 100 ... 240 V AC / 50 / 60 Hz
Output: 5 V / 1 A DC

Operating time of the accumulator

Approximately 18 h: Simulation and measurement mode with low LCD lighting
Approximately 8 h: 12 mA (24 V) measurement mode with low LCD lighting

Dimension

Other, of a width of not more than 600 mm

Weight

Approximately 300 g

Degree of protection

The following information is provided:

Operating conditions

Accumulator: 0 ... +55 °C / 30 ... 90 % H.r.
Electricity grid: 0 ... +45 °C / 30 ... 90 % H.r.

Storage conditions

-20 ... +60 °C / 30 ... 90% H.r. without condensation

Warming time

About 15 minutes.