Radiómetro PCE-MFM 3500
- Code / Model:
- Manufacturer: PCE Instruments
- Department:
- Category of products:
- Sub-category: Instruments and equipment
- You've seen it before.: 1
Radiometer with external sensor / Range: up to 30,000 Gauss /
Level bubble integrated into the sensor / Internal memory for 7000 values /
Software for real-time memory reading and visualization / Cable length: 1 m /
Battery power and electricity grid / Battery operating time: approximately 100 hours
Level bubble integrated into the sensor / Internal memory for 7000 values /
Software for real-time memory reading and visualization / Cable length: 1 m /
Battery power and electricity grid / Battery operating time: approximately 100 hours
The radiometer is a device for measuring magnetic fields. This radiometer is suitable for measuring static and dynamic magnetic fields. The flat tip of the external sensor allows the user to measure even in very small holes. The radiometer's measuring range reaches up to 30,000 Ga
The radiometer also integrates the data recording function. Thanks to the radiometer's internal memory it can store measurement values automatically or manually. To analyze the recorded data you simply need to connect the radiometer to a computer via the mini USB interface. The software allows the user to
The radiometer also integrates the data recording function. Thanks to the radiometer's internal memory it can store measurement values automatically or manually. To analyze the recorded data you simply need to connect the radiometer to a computer via the mini USB interface. The software allows the user to
- Can measure the DC and AC magnetic fields (40~500Hz)
- Adjustable alarm function
- Data memory with PC connection
- Temperature measurement
- AC/DC measurement switched
- Sensor integrates level bubble
- Rubber case for handheld equipment
- Adjustable alarm function
- Data memory with PC connection
- Temperature measurement
- AC/DC measurement switched
- Sensor integrates level bubble
- Rubber case for handheld equipment
| Function | Rate | Resolution adopted by the Council | Accuracy |
| The DC Militia | Other, not further worked than hot-rolled | 0,01 mT | ±(4 % + 10 digits) |
| Other | 0,1 mT | ±(4 % + 10 digits) | |
| 2000,1 ... 3000,0 mT | 0,1 mT | N/A | |
| DC Gauss | Other, of a kind used for the manufacture of goods | 0,1 G | ±(4 % + 10 digits) |
| 20,000 g | 1 G | ±(4 % + 10 digits) | |
| 20,0001 ... 30,000 G | 1 G | N/A | |
| The Commission has also adopted a proposal for a regulation on the | Other, not further worked than hot-rolled | 0,01 mT | ±(5 % + 20 digits) |
| Other, not further worked than hot-rolled | 0,1 mT | ±(5 % + 20 digits) | |
| The Commission shall adopt implementing acts in accordance with Article 21 of this Regulation. | Other, not elsewhere specified or included | 0,1 G | ±(5 % + 20 digits) |
| 15,000 g | 1 G | ±(5 % + 20 digits) | |
| The precision is subject to the following environmental conditions: 23 °C ±5 °C, <80 % H.r. | |||
| The temperature | -20 ... 50 °C -4 ... 122 °F | The following conditions shall apply: The following is the list of the following: | ±1,0 °C The following conditions shall apply: |
| Memory | Manual: up to 200 values Automatic: up to 7000 values | ||
| Interface | Mini USB with a USB | ||
| Screen | LCD, five digits, up to 30000 | ||
| Sensor | External with plug, cable length: approximately 1 m | ||
| Automatically disconnected | After 15 minutes or the function is off | ||
| Battery feed | 6 x 1,5 V, type AAA | ||
| Operating time with batteries | About 100 hours | ||
| Power by network cable | Input: 100 to 240 V Output: 9 V, 0.5 A | ||
| Operating conditions | 5 ... 40 °C / 41 ... 104 °F, < 80% H.r. | ||
| Storage conditions | 0 ... 50 °C, < 70% H.r. | ||
| Height above sea level | Maximum 2000 m | ||
| Degree of pollution | 2 | ||
| Dimension | Other, of a thickness of not more than 15 mm | ||
| Weight | Approximately 320 g (without batteries) | ||