Medidor de automoción PCE-CTI 10
- Code / Model:
- Manufacturer: PCE Instruments
- Department:
- Category of products:
- Sub-category: Instruments and equipment
- You've seen it before.: 1
Other electrical equipment, including electrical equipment
Expandable up to 3000 A AC with flexible current grip / VFD mode /
The power consumption shall be measured at the level of the power consumption.
Expandable up to 3000 A AC with flexible current grip / VFD mode /
The power consumption shall be measured at the level of the power consumption.
The automotive meter is the ideal device for measuring voltages up to 1500 V AC/DC and currents up to 1000 A AC/DC. This automotive meter measures the current inductively. For this, you simply have to place the line or cable inside the automotive meter. This has the
In addition to measuring the voltage and current, the automotive meter also measures the start current. The start current is the current that is especially high at the time of start-up. These currents are generated when a motor starts, such as ventilation systems or stadium lighting. This measur
The car meter integrates a Bluetooth 4.0 interface, which allows you to connect the car meter to an Android or iOS terminal.
The car meter allows reading of each individual measurement point. The car meter memory has 16 groups and a total memory for 100 000 values. The data is stored in the car meter with date and time.
The additional current gauge is flexible and therefore can be adapted to the measuring task. With the optional current gauge it can perform current measurements of up to 3000 A AC. The flexible current gauge is connected to the voltage input of the automotive gauge. The automotive ga
The automotive meter is used for many functions and a high measurement range, for example for checking photovoltaic systems, hydraulic and wind power plants or for installing fixed electric machines.
In addition to measuring the voltage and current, the automotive meter also measures the start current. The start current is the current that is especially high at the time of start-up. These currents are generated when a motor starts, such as ventilation systems or stadium lighting. This measur
The car meter integrates a Bluetooth 4.0 interface, which allows you to connect the car meter to an Android or iOS terminal.
The car meter allows reading of each individual measurement point. The car meter memory has 16 groups and a total memory for 100 000 values. The data is stored in the car meter with date and time.
The additional current gauge is flexible and therefore can be adapted to the measuring task. With the optional current gauge it can perform current measurements of up to 3000 A AC. The flexible current gauge is connected to the voltage input of the automotive gauge. The automotive ga
The automotive meter is used for many functions and a high measurement range, for example for checking photovoltaic systems, hydraulic and wind power plants or for installing fixed electric machines.
- Free app for Android and iOS
- Range: 0 ... 1500 V AC/DC
- Memory for 100 000 values
- Measurement of the starting current
- Additional: LowZ voltage measurement
- TFT screen of 2.36 "
- Voltage and temperature measurement
- Optional: calibration certificate
- Range: 0 ... 1500 V AC/DC
- Memory for 100 000 values
- Measurement of the starting current
- Additional: LowZ voltage measurement
- TFT screen of 2.36 "
- Voltage and temperature measurement
- Optional: calibration certificate
| Continuous voltage | ||
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 0,1 mV | |
| Accuracy | ±(0,8% of the value + 8 digits) | |
| Rate | ±6 000 V | |
| Resolution adopted by the Council | 0,001 V | |
| Accuracy | ±(0,5% of the value + 5 digits) | |
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 0,01 V | |
| Accuracy | ±(0,5% of the value + 5 digits) | |
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 0,1 V | |
| Accuracy | ±(0,8% of the value + 5 digits) | |
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 1 V | |
| Accuracy | ±(0,8% of the value + 5 digits) | |
| The following shall be added to the list of the components of the vehicle: | ||
| Alternating voltage | ||
| Rate | 0,000 ... 6,000 V | |
| Resolution adopted by the Council | 0,001 V | |
| Accuracy | 50 ... 60 Hz: ±(1,2% of the value + 5 digits) 61 ... 1 kHz: ±(2,5% of the value + 5 digits) | |
| Rate | 0,00 ... 60,00 V | |
| Resolution adopted by the Council | 0,01 V | |
| Accuracy | 50 ... 60 Hz: ±(1,2% of the value + 5 digits) 61 ... 1 kHz: ±(2,5% of the value + 5 digits) | |
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 0,1 V | |
| Accuracy | 50 ... 60 Hz: ±(1,2% of the value + 5 digits) 61 ... 1 kHz: ±(2,5% of the value + 5 digits) | |
| Rate | 0 ... 1500 V | |
| Resolution adopted by the Council | 1 V | |
| Accuracy | 50 ... 60 Hz: ±(1,2% of the value + 5 digits) 61 ... 1 kHz: ±(2,5% of the value + 5 digits) | |
| The accuracy is applicable between 10 ... 100 % of the respective measurement range and a sinusoidal wave The following shall be added to the list of the following: The accuracy of the 'PEAK' function is ±10 % of the value The response time of the "PEAK" function is 1 ms | ||
| Alternating voltage with a small input impedance (LowZ) | ||
| Rate | 0,000 ... 6,000 V | |
| Resolution adopted by the Council | 0,001 V | |
| Accuracy | ±(3,0% of the value + 40 digits) | |
| Rate | 0,00 ... 60,00 V | |
| Resolution adopted by the Council | 0,01 V | |
| Accuracy | ±(3,0% of the value + 40 digits) | |
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 0,1 V | |
| Accuracy | ±(3,0% of the value + 40 digits) | |
| The following shall be added to the list of the relevant technical specifications: The accuracy is applicable between 10 ... 100 % of the respective measurement range and a sinusoidal wave | ||
| Continuous and alternating voltage (50 ... 1 kHz) | ||
| Rate | 0,000 ... 6,000 V | |
| Resolution adopted by the Council | 0,001 V | |
| Accuracy | ±(2,5% of the value + 40 digits) | |
| Rate | 0,00 ... 60,00 V | |
| Resolution adopted by the Council | 0,01 V | |
| Accuracy | ±(2,5% of the value + 40 digits) | |
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 0,1 V | |
| Accuracy | ±(2,5% of the value + 40 digits) | |
| Rate | 0 ... 1000 V | |
| Resolution adopted by the Council | 1 V | |
| Accuracy | ±(2,5% of the value + 40 digits) | |
| The following shall be added to the list of the relevant technical specifications: | ||
| Continuous and alternating voltage with a small input impedance (LowZ) | ||
| Rate | 0,000 ... 6,000 V | |
| Resolution adopted by the Council | 0,001 V | |
| Accuracy | ±(3,5% of the value + 40 digits) | |
| Rate | 0,00 ... 60,00 V | |
| Resolution adopted by the Council | 0,01 V | |
| Accuracy | ±(3,5% of the value + 40 digits) | |
| Rate | The following conditions shall apply: | |
| Resolution adopted by the Council | 0,1 V | |
| Accuracy | ±(3,5% of the value + 40 digits) | |
| The following shall be added to the list of the relevant technical specifications: | ||
| The following shall be added to the list of the following: | ||
| Rate | 100 ... 1000 AC/DC | |
| Continuous current | ||
| Rate | Other, not further worked than cutting | |
| Resolution adopted by the Council | 0,01 A | |
| Accuracy | ±(3,0% of the value + 8 digits) | |
| Rate | 600,0 A | |
| Resolution adopted by the Council | 0,1 A | |
| Accuracy | ±(3,0% of the value + 8 digits) | |
| Rate | 1000 A | |
| Resolution adopted by the Council | 1 A | |
| Accuracy | ±(3,0% of the value + 8 digits) | |
| The following conditions shall apply: | ||
| Continuous current (50 ... 60 Hz) | ||
| Rate | 0,00 ... 60,00 A | |
| Resolution adopted by the Council | 0,01 A | |
| Accuracy | ±(2,5% of the value + 5 digits) | |
| Rate | 0,0 ... 600,0 A | |
| Resolution adopted by the Council | 0,1 A | |
| Accuracy | ±(2,5% of the value + 5 digits) | |
| Rate | 0 ... 1000 A | |
| Resolution adopted by the Council | 1 A | |
| Accuracy | ±(2,5% of the value + 5 digits) | |
| The following conditions shall apply: The accuracy is applicable between 10 ... 100 % of the respective measurement range and a sinusoidal wave The start current only serves as a reference | ||
| The following shall be added to the list of the relevant technical provisions: | ||
| Rate | 0,00 ... 30,00 A | |
| Resolution adopted by the Council | 0,01 A | |
| Accuracy | ±(3,0% of the value + 5 digits) | |
| Rate | 0,0 ... 300,0 A | |
| Resolution adopted by the Council | 0,1 A | |
| Accuracy | ±(3,0% of the value + 5 digits) | |
| Rate | 0 ... 3000 A | |
| Resolution adopted by the Council | 1 A | |
| Accuracy | ±(3,0% of the value + 5 digits) | |
| Frequency range | 50 ... 400 Hz | |
| The following shall be added: | The following is the list of the products used: | |
| The following conditions shall apply: The accuracy is applicable between 10 ... 100 % of the respective measurement range and a sinusoidal wave | ||
| Resistance and continuity test | ||
| Rate | 0,0 ... 600,0 Ω | |
| Resolution adopted by the Council | 0,1 Ω | |
| Accuracy | ±(1,0% of the value + 10 digits) | |
| Rate | 0,000 ... 6,000 kΩ | |
| Resolution adopted by the Council | 0,001 kΩ | |
| Accuracy | ±(0,8% of the value + 5 digits) | |
| Rate | 0,00 ... 60,00 kΩ | |
| Resolution adopted by the Council | 0,01 kΩ | |
| Accuracy | ±(0,8% of the value + 5 digits) | |
| Rate | 0,0 ... 600,0 kΩ | |
| Resolution adopted by the Council | 0,1 kΩ | |
| Accuracy | ±(0,8% of the value + 5 digits) | |
| Rate | 0,000 ... 6,000 MΩ | |
| Resolution adopted by the Council | 0,001 MΩ | |
| Accuracy | ±(0,8% of the value + 5 digits) | |
| Rate | 0,00 ... 60,00 MΩ | |
| Resolution adopted by the Council | 0,01 MΩ | |
| Accuracy | ±(2,5% of the value + 10 digits) | |
| Acoustic signal in continuity test: < 50 Ω The following conditions shall apply: | ||
| Frequency | ||
| Rate | 0,00 ... 60,00 Hz | |
| Resolution adopted by the Council | 0,01 Hz | |
| Accuracy | ±(0,2% of the value + 5 digits) | |
| Rate | The frequency of the sound is not less than 0,0 ... 600,0 Hz | |
| Resolution adopted by the Council | 0,1 Hz | |
| Accuracy | ±(0,2% of the value + 5 digits) | |
| Rate | 0,000 ... 6,000 kHz | |
| Resolution adopted by the Council | Other, of a kind used for the manufacture of goods | |
| Accuracy | ±(0,2% of the value + 5 digits) | |
| Rate | The following is the list of the parameters used: | |
| Resolution adopted by the Council | Other, of a kind used for the manufacture of goods | |
| Accuracy | ±(0,2% of the value + 5 digits) | |
| Rate | The frequency of the transmission shall be: | |
| Resolution adopted by the Council | Other, not further worked than hot-rolled | |
| Accuracy | ±(0,2% of the value + 5 digits) | |
| Rate | 0,000 ... 6,000 MHz | |
| Resolution adopted by the Council | 0,001 MHz | |
| Accuracy | ±(0,2% of the value + 5 digits) | |
| The following conditions shall apply: Sensitivity: > 2 Vrms with a working cycle between 20 ... 80 % and a frequency < 100 kHz > 5 Vrms with a working cycle between 20 ... 80 % and a frequency > 100 kHz | ||
| The following is the list of the activities of the Agency: | ||
| Rate | The Commission shall adopt delegated acts in accordance with Article 21 of this Regulation. | |
| Resolution adopted by the Council | 0,1% | |
| Accuracy | ±(1,2% of the value + 8 digits) | |
| Frequency range: 40 ... 10 kHz; Pulse amplitude ±5 V with a duration between 100 μs ... 100 mS | ||
| Capacity | ||
| Accuracy | 0,00 ... 60,00 nF | |
| Rate | 0,01 nF | |
| Resolution adopted by the Council | ±(3,0% of the value + 20 digits) | |
| Accuracy | Other, of a kind used for the manufacture of goods | |
| Rate | 0,1 nF | |
| Resolution adopted by the Council | ±(3,0% of the value + 8 digits) | |
| Accuracy | 0,000 to 6,000 μF | |
| Rate | 0,001 μF | |
| Resolution adopted by the Council | ±(3,0% of the value + 8 digits) | |
| Accuracy | 0,00 ... 60,00 μF | |
| Rate | 0,01 μF | |
| Resolution adopted by the Council | ±(3,0% of the value + 8 digits) | |
| Accuracy | Other, of a kind used for the manufacture of goods | |
| Rate | 0,1 nF | |
| Resolution adopted by the Council | ±(3,0% of the value + 8 digits) | |
| Accuracy | 0 ... 6000 μF | |
| Rate | 1 μF | |
| Resolution adopted by the Council | ±(3,5% of the value + 20 digits) | |
| Accuracy | 0,00 ... 60,00 mF | |
| Rate | 0,01 mF | |
| Resolution adopted by the Council | only as a reference | |
| Accuracy | 0,0 ... 600,0 mF | |
| Rate | 0,1 mF | |
| Resolution adopted by the Council | only as a reference | |
| The following conditions shall apply: | ||
| Temperature with K-type thermoelectricity | ||
| Rate | -40 ... 600,0 °C | |
| Resolution adopted by the Council | The following conditions shall apply: | |
| Accuracy | ±(1,5% of the value + 3 °C) | |
| Rate | 600 ... 1000 °C | |
| Resolution adopted by the Council | 1 °C | |
| Accuracy | ±(1,5% of the value + 3 °C) | |
| Rate | -40 ... 600,0 °F | |
| Resolution adopted by the Council | 0,1 °F | |
| Accuracy | ±(1,5% of the value + 5,4 °F) | |
| Rate | 600 to 1800 °F | |
| Resolution adopted by the Council | 1 °F | |
| Accuracy | ±(1,5% of the value + 5,4 °F) | |
| Rate | The following is the list of the products concerned: | |
| Resolution adopted by the Council | 0,1 K | |
| Accuracy | ±(1,5% of the value + 3 K) | |
| Rate | 600 ... 1273 K | |
| Resolution adopted by the Council | 1 K | |
| Accuracy | ±(1,5% of the value + 3 K) | |
| The following conditions shall apply: | ||
| Test of diodes | ||
| Test current | Other, of a kind used for the manufacture of goods | |
| Maximum voltage | 3,3 VDC | |
| Other specifications | ||
| Memory | 16 groups, total memory of 100 000 values | |
| The following is the list of the types of products used: | Other, of a width of not more than 30 mm | |
| The following is the list of the types of products used: | Other, of a thickness of not more than 30 mm | |
| Other, of a kind used for the manufacture of goods | Other, of a width of not more than 30 mm | |
| Protection | Other, not further worked than cutting | |
| Interface | Bluetooth 4.0 and the | |
| Degree of pollution | 2 | |
| Category of the product | The following conditions shall apply: | |
| Maximum working height | 2000 m / 6562 ft | |
| Other, of a kind used for the manufacture of goods | Lithium, 7,4 V, 1200 mAh | |
| Loading food | Input Output | 100 ... 240 V AC, 50 ... 60 Hz 12 V DC, 2 A |
| Type of charger socket | The Commission has also adopted a number of measures to combat the spread of the virus. | |
| Indication of the state of the battery | Yes, I did. | |
| Automatically disconnected | Deactivated, 15, 30 or 60 minutes | |
| Screen | The following table shows the number of samples taken: | |
| Updates on the screen | 3 Hz | |
| Reference conditions | The following conditions shall apply: < 80% H.r. without condensation | |
| Operating conditions | 5 ... 40 °C / 41 ... 104 °F < 80% H.r. without condensation | |
| Storage conditions | -20 ... 60 °C / -4 ... 140 °F < 80% H.r. without condensation | |
| Dimension | Other, of a thickness of not more than 0,75 mm | |
| Weight | 481 g | |